Section 14: SCPI command reference Model 6517B Electrometer Reference Manual
14-18 6517B-901-01 Rev. C / August 2015
Command Description Default parameter SCPI
:RSWitch?
Query switch setting (surface or volume)
:USER
Path to configure user test fixture:
:RSELect
<name>
Specify measurement type (SURFace or
VOLume)
SURFace
:RSELect?
Query measurement type
:KSURface
<NRf>
Surface; specify Ks (0.001 to 999.999) 1.000
:KSURface?
Query, Ks
:KVOLume <NRf>
Volume; specify Kv (0.001 to 999.999) 1.000
:KVOLume?
Query Kv
:CHARge
Path to configure Coulombs

:APERture <n>
Set integration rate in seconds (166.67e-6
to 200e-3)
(Note 1)

:AUTO <b>
Enable or disable auto aperture OFF
:AUTO ONCE
Enable and then disable auto aperture
:AUTO?
Query auto aperture
:APERture?
Query aperture (integration rate) 1

:NPLCycles <n>
Set integration rate (line cycles; 0.01 to
10)
OFF

:AUTO <b>
Enable or disable auto NPLC
:AUTO ONCE
Enable and then disable auto NPLC
:AUTO?
Query auto line cycle integration
:NPLCycles?
Query line cycle integration rate

:CHARge

:RANGe
Path to configure measurement range:

[:UPPer] <n>
Select range (0 to +2.1e-6) 2e-6

[:UPPer]?
Query range

:AUTO <b>
Enable or disable auto range ON

:AUTO ONCE
Set range based on present input signal

:LGRoup <name>
Specify HIGH or LOW auto range limit HIGH
:LGRoup?
Query auto range limit
:AUTO?
Query auto range
:REFerence <n>
Specify reference (-2.1e-6 to +2.1e-6) 0

:STATe <b>
Enable or disable reference OFF

:STATe?
Query state of reference

:ACQuire
Use input signal as reference
:REFerence?
Query reference value

:DIGits <n>
Specify measurement resolution (4 to 7) 6
:AUTO <b>
Enable or disable auto resolution
:AUTO ONCE
Enable and then disable auto resolution
:AUTO?
Query auto resolution
:DIGits?
Query resolution
:AVERage
Path to control the average filter

Table of Contents

Other manuals for Keithley 6517B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6517B and is the answer not in the manual?

Keithley 6517B Specifications

General IconGeneral
BrandKeithley
Model6517B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals