Model 6517B Electrometer Reference Manual Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015 14-19
Command Description Default parameter SCPI
:TYPE <name>
Select filter type (NONE, SCALar,
ADVanced)
NONE
:TYPE?
Query filter type
:TCONtrol
<name>
Select filter type: (MOVing or REPeat)
MOVing
:TCONtrol?
Query filter type
:COUNt <n>
Specify filter count (1 to 100) 10
:COUNt?
Query filter count
:ADVanced
Path to configure and control advanced
filter
:NTOLerance
<n>
Specify noise tolerance level (0 to 100 %) 1
:NTOLerance?
Query noise tolerance level
[:STATe] <b>
Enable or disable filter OFF
[:STATe]?
Query state of digital filter
:MEDian
Path to control the median filter:
[:STATe] <b>
Enable or disable median filter ON
[:STATe]?
Query state of median filter
:RANK <NRf>
Specify rank for median filter 1
:RANK?
Query rank
:ADIScharge
Path to control auto discharge:
[:STATe] <b>
Enable or disable auto discharge OFF
[:STATe]?
Query state of auto discharge
:LEVel <NRf>
Specify level (-2.2e6 to 2.2e6) 2e-6
:LEVel?
Query level
NOTES:
1. 16.67 ms for 60 Hz power line frequency and 20ms for 50 Hz line frequency.
2. *RST — OFF, :SYSTem:PRESet — ON.
3. *RST — REPeat, :SYSTem:PRESet — MOVing.
4. *RST — ON, :SYSTem:PRESet — OFF.