EasyManua.ls Logo

Keithley 6517B

Keithley 6517B
498 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Section 14: SCPI command reference Model 6517B Electrometer Reference Manual
14-30 6517B-901-01 Rev. C / August 2015
Command Description Default parameter SCPI
:SVOLtage[1]
<NRf>
Specify source voltage 1; -1000 to
1000
+1 V
:SVOLtage[1]?
Query source voltage
:MDELay[1]
<NRf>
Specify measure delay 1; 0 to 99999.9
(sec)
1 sec
:MDELay[1]?
Query measure delay
:SVOLtage2
<NRf>
Specify source voltage 2; -1000 to
1000
+2 V
:SVOLtage2?
Query source voltage
:MDELay2 <NRf>
Specify measure delay 2; 0 to 99999.9
(sec)
1 sec
:MDELay2?
Query measure delay
:SRESistivity
Surface resistivity test path:
:PDTime <NRf>
Specify pre-discharge time; 0 to
9999.9 (sec)
0.2 sec
:PDTime?
Query pre-discharge time
:SVOLtage
<NRf>
Specify bias voltage; -1000 to 1000 +500 V
:SVOLtage?
Query bias voltage
:STIME <NRf>
Specify bias time; 0 to 99999.9 (sec) 1 sec
:STIME?
Query bias time
:MVOLtage
<NRf>
Specify measure voltage; -1000 to
1000
+500 V
:MVOLtage?
Query measure voltage
:MTIMe <NRf>
Specify measure time; 0 to 9999.9
(sec)
1 sec
:MTIMe?
Query measure time
:DTIMe <NRf>
Specify discharge time; 0 to 99999.9
(sec)
2 sec
:DTIMe?
Query discharge time
:VRESistivity
Volume resistivity test path:
:PDTime <NRf>
Specify pre-discharge time; 0 to
99999.9 (sec)
10 sec
:PDTime?
Query pre-discharge time
:SVOLtage
<NRf>
Specify bias voltage; -1000 to 1000 +500 V
:SVOLtage?
Query bias voltage
:STIME <NRf>
Specify bias time; 0 to 99999.9 (sec) 1 sec
:STIME?
Query bias time
:MVOLtage
<NRf>
Specify measure voltage; -1000 to
1000
+500 V
:MVOLtage?
Query measure voltage
:MTIMe <NRf>
Specify measure time; 0 to 9999.9
(sec)
1 sec
:MTIMe?
Query measure time
:DTIMe <NRf>
Specify discharge time; 0 to 99999.9
(sec)
2 sec
:DTIMe?
Query discharge time
:TSEQuence

Table of Contents