Model 6517B Electrometer Reference Manual Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015 14-31
Command Description Default parameter SCPI
:ALTPolarity
Alternating polarity resistance/resistivity
test path:
:OFSVoltage
<NRf>
Specify offset voltage; -1000 to 1000 0 V
:OFSVoltage?
Query offset voltage
:ALTVoltage
<NRf>
Specify alternating voltage; -1000 to
1000
10 V
:ALTVoltage?
Query alternating voltage
:MTIMe <NRf>
Specify measurement time; 0.5 to
9999.9 (sec)
15 sec
:MTIMe?
Query measurement time
:DISCard <NRf>
Specify number of initial readings to
discard;
0 to 9999
3
:DISCard?
Query number of initial readings to
discard
:READings
<NRf>
Specify number of readings to store 1
:READings?
Query number of readings to store
:SIResistance
Surface insulation resistance test path:
:SVOLtage
<NRf>
Specify bias voltage; -1000 to 100 +50 V
:SVOLtage?
Query bias voltage
:STIME <NRf>
Specify bias time; 0 to 99999.9 (sec) 1 sec
:STIME?
Query bias time
:MVOLtage
<NRf>
Specify measure voltage; -1000 to
1000
+100 V
:MVOLtage?
Query measure voltage
:MTIMe <NRf>
Specify measure time; 0 to 9999.9
(sec)
1 sec
:MTIMe?
Query measure time
:SQSWeep
Square wave sweep test path:
:HLEVel <NRf>
Specify high-level voltage; -1000 to
1000
+1 V
:HLEVel?
Query high-level voltage
:HTIMe <NRf>
Specify high-level time; 0 to 9999.9
(sec)
1 sec
:HTIMe?
Query high-level time
:LLEVel <NRf>
Specify low-level voltage; -1000 to
1000
-1 V
:LLEVel?
Query low-level voltage
:LTIMe <NRf>
Specify low-level time; 0 to 9999.9
(sec)
1 sec
:LTIMe?
Query low-level time
:COUNt <NRf>
Specify number cycles
:COUNt?
Query cycle count
:STSWeep
Staircase sweep test path:
:STARt <NRf>
Specify start voltage; -1000 to 1000 +1 V
:STARt?
Query start voltage
:STOP <NRf>
Specify stop voltage; -1000 to 1000 +10 V