Model 6517B Electrometer Reference Manual Section 15: Calibration procedure
6517B-901-01 Rev. C / August 2015 15-27
Error Code Description
+321 Buffer sizing error; set to MAX
+322 Buffer sizing error; set to MIN
+350 2 V offset out of specification
+351 2 V positive gain out of specification
+352 2 V negative gain out of specification
+353 20 V offset out of specification
+354 20 V positive gain out of specification
+355 20 V negative gain out of specification
+356 200 V offset out of specification
+357 200 V positive gain out of specification
+358 200 V negative gain out of specification
+359 20 pA offset out of specification
+360 20 pA positive gain out of specification
+361 20 pA negative gain out of specification
+362 200 pA offset out of specification
+363 200 pA positive gain out of specification
+364 200 pA negative gain out of specification
+365 2 nA offset out of specification
+366 2 nA positive gain out of specification
+367 2 nA negative gain out of specification
+368 20 nA offset out of specification
+369 20 nA positive gain out of specification
+370 20 nA negative gain out of specification
+371 200 nA offset out of specification
+372 200 nA positive gain out of specification
+373 200 nA negative gain out of specification
+374 2 mA offset out of specification