3-44 Keysight E5260/E5270 User’s Guide, Edition 6
Installation
Maintenance
Self-test and Diagnostics
Keysight E5260/E5270 provides the following functions to check the operation.
Perform the following functions as necessary.
Self-test
Refer to “To Perform Self-test” on page 5-7.
Self-calibration
Refer to “To Perform Self-Calibration” on page 5-8.
Diagnostics
Refer to “To Perform Diagnostics” on page 5-9.
WARNING Interlock Open/Close test and High Voltage LED test
Interlock function limits the maximum output voltage to ± 42 V when the
measurement terminal is touchable. For safety, this function must be checked
before using the E5260/E5270 after power on at least once a day by performing
the diagnosis.
Test d'ouverture/de fermeture Interlock et test High Voltage LED
La fonction de verrouillage limite la tension de sortie maximale à ± 42 V
lorsque la borne de mesure peut être touchée. Pour des raisons de sécurité,
cette fonction doit être vérifiée avant d'utiliser le E5260/E5270 après la mise
sous tension, au moins une fois par jour en effectuant le diagnostic.

Table of Contents

Other manuals for Keysight Technologies E5260 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E5260 Series and is the answer not in the manual?

Keysight Technologies E5260 Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelE5260 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals