Keysight E5260/E5270 User’s Guide, Edition 6 5-7
Front Panel Operations
Setting Up Keysight E5260/E5270
To Perform Self-test
1. Open the measurement terminals of the Keysight E5260/E5270.
2. Press the
Shift
key and the
Menu
key.
3. Move the cursor to SELFTEST, then press the
Enter
key.
4. Move the cursor to EXECUTE, then press the
Enter
key.
5. Press the arrow keys to select the self-test item; ALL, FRAME, or Slot n (n: 1 or
2 for 2-ch mainframe, 1 to 8 for 8-ch mainframe).
6. Press the
Enter
key to start the self-test, or press the
Exit
key to cancel it.
During the self-test, press the
Exit
key to abort the self-test.
7. Press the
Exit
key three times to close the setup menu.
To display test results
1. Press the
Shift
key and the
Menu
key.
2. Move the cursor to SELFTEST, then press the
Enter
key.
3. Move the cursor to RESULT, then press the
Enter
key.
4. Press the arrow keys to select the self-test item; FRAME or Slot n (n: 1 or 2 for
2-ch mainframe, 1 to 8 for 8-ch mainframe). The test result is displayed for each
item.
5. Press the
Exit
key three times to close the setup menu.
To use modules that failed self-test
The following procedure enables you to use a module that failed self-test or
self-calibration.
1. Press the
Shift
key and the
Menu
key.
2. Move the cursor to SELFTEST, then press the
Enter
key.
3. Move the cursor to RECOVER, then press the
Enter
key.
4. Press the arrow keys to select YES for the message Recover Module:.
5. Press the
Enter
key to perform this operation, or press the
Exit
key to cancel the
operation.
6. Press the
Exit
key twice to close the setup menu.