2-8 Keysight E5260/E5270 User’s Guide, Edition 6
Introduction
Measurement Modules
Measurement Modules
The source/monitor unit (SMU) can force a constant voltage or current, and can
measure a DC current or voltage.
Figure 2-1 is a simplified SMU circuit diagram. The SMU can perform the
following operations:
Apply voltage and measures current
Apply current and measures voltage
Apply voltage and measures voltage
Apply current and measures current
NOTE To select the SMU operation mode in the GPIB remote condition, enter the CMM
command.
Figure 2-1 Simplified SMU Circuit Diagram
Compliance
The SMU has a compliance feature that limits output voltage or current to prevent
damage to the device under test. When the SMU applies voltage, you can specify
current compliance. When the SMU applies current, you can specify voltage
compliance. For details about the compliance, refer to “Compliance” on page 6-9.

Table of Contents

Other manuals for Keysight Technologies E5260 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E5260 Series and is the answer not in the manual?

Keysight Technologies E5260 Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelE5260 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals