Keysight E5260/E5270 User’s Guide, Edition 6 7-11
If You Have a Problem
Error Codes
Error Codes
If errors occur, error codes are stored in the error buffer. To read the error code,
execute the “ERR?” command. To read the error message, execute the “EMG?”
command.
The output of the error codes is in the order that they occurred, and the first four
error codes are stored in the buffer. If no errors occurred, ”0, 0, 0, 0” is returned.
Operation Error
100 Undefined GPIB command.
Send the correct command.
102 Incorrect numeric data syntax.
Correct the data syntax.
103 Incorrect terminator position.
Correct the command syntax. The number of parameters will be
incorrect.
120 Incorrect parameter value.
Correct the parameter value.
121 Channel number must be 1 to 2, or 1 to 8.
Correct the channel number. The channel number must be 1 to 2 for the
Keysight E5262A/E5263A, or 1 to 8 for the Keysight
E5260A/E5270B.
122 Number of channels must be corrected.
Check the MM, FL, CN, CL, IN, DZ, or RZ command, and correct the
number of channels.
123 Compliance must be set correctly.
Incorrect compliance value was set. Set the compliance value correctly.
124 Incorrect range value for this channel.
Check the range value available for the channel, and correct the range
value.