3-36 Keysight E5260/E5270 User’s Guide, Edition 6
Installation
Connecting Measurement Devices
Using Test Fixture
1. Press the
OutCh
and
On/Off
keys to set the output off.
2. Select a proper socket module for your DUT, then set the socket module on the
test fixture.
3. Mount your DUT on the socket module.
4. Connect between the socket module and the test fixture by using the proper test
leads.
5. Close the fixture cover.
To force more than ± 42 V, close the fixture cover. Otherwise, the interlock
function will stop the Keysight E5260/E5270 output.
To connect between the socket module and the test fixture, you can use test leads
that have the following terminals:
• Miniature banana-to-miniature banana
• Miniature banana-to-pin plug
• Miniature banana-to-miniature clip
WARNING Set the instrument output off before connecting or disconnecting test leads.
Press the front panel
OutCh
and
On/Off
keys to set the output off. And confirm
that the front panel HIGH VOLTAGE indicator is not lit.
Désactivez la sortie de l'appareil avant de brancher ou de débrancher cordons
de mesure.
Appuyez sur les touche
OutCh
et
On/Off
du panneau avant du pour désactiver la
sortie. Et confirmez que l'indicateur HIGH VOLTAGE du panneau avant du
n'est pas allumé.
WARNING Make sure that the fixture cover is closed properly before starting
measurement. Do not perform the measurement when a test lead is protruding
from the test fixture and the cover.
Assurez-vous que le couvercle est fermé correctement avant de commencer la
mesure. Ne pas effectuer la mesure lorsqu’un cordon de test dépasse du
couvercle de l'appareil.
CAUTION Do not touch the terminals of the test leads. Oil, perspiration, and dirt prevent good
electrical contact, deteriorate insulation, and degrade measurement accuracy.