Keysight E5260/E5270 User’s Guide, Edition 6 3-5
Installation
Requirements
Installation Requirements
WARNING Do not operate the instrument in dusty environment, or in the presence of
flammable gasses, corrosive gasses, or fumes.
Ne pas utiliser l’appareil dans un endroit poussiéreux, ou en présence de gaz
inflammables, corrosifs ou de fumée.
• Environmental conditions for this instrument are documented in “Operating
Environment” on page 3-4. In principle, this instrument should only be operated
indoors in a controlled environment.
• Install the instrument horizontal place then face the roof up.
• Fan cools the instrument by drawing air through the sides and exhausting it out
the back. The instrument must be installed in a location that allows sufficient
space at the sides and back of the instrument for adequate air circulation.
If the airflow is restricted, the internal operating temperature will be higher. This
may reduce the instrument’s reliability, or cause the thermal-protection circuits
to turn the instrument off.
• Make sure that there is enough space for connecting/disconnecting the power
cable easily.
Power Cable
WARNING FIRE HAZARD: Use only the power cable supplied with your instrument.
Using other types of power cable may cause overheating of the power cable,
resulting in fire.
SHOCK HAZARD: The power cable provides the chassis ground through a
third conductor. Be sure to connect to a three-conductor type power outlet with
the correct pin grounded.
RISQUE D’INCENDIE : utilisez uniquement le câble d'alimentation fourni
avec votre appareil. L’utilisation d’autres types de câble d’alimentation peut
provoquer une surchauffe du câble d’alimentation et provoquer un incendie.
RISQUE DE CHOC ÉLECTRIQUE: le câble d’alimentation fournit la masse
du châssis par le biais d’un troisième conducteur. Assurez-vous de connecter la
prise d'alimentation de type trois conducteurs avec la broche correcte mise à la
terre.