3-10 Keysight E5260/E5270 User’s Guide, Edition 6
Installation
Inspection and Installation
To Remove or Install Plug-in Module
This section provides the plug-in module removal and installation instructions.
WARNING To prevent electrical shock, turn off the mainframe and remove the power
cable before starting the instruction.
Afin d'éviter toute décharge électrique, mettez l'appareil hors tension et
débranchez le cordon d'alimentation avant de commencer l'instruction.
CAUTION Be careful about the interconnection connector and use clean handling and
anti-static procedures when unpacking or doing the following procedure to prevent
damage.
CAUTION To prevent instrument damage, cover over the empty slot by installing the blank
panel instead of the plug-in module.
NOTE Plug-in module is shipped from the factory after the specifications are confirmed.
Keysight can guarantee that the modules will function and that the module
performance is designed to meet its specifications. However, if you have any
trouble, contact your nearest Keysight Technologies service center.
If you require that the modules be guaranteed to their specifications (for ISO
compliance, etc.), then you must perform a calibration on the instrument (modules
and mainframe together). This requirement arises because the mainframes contain
both a Ground Unit and an A/D converter, which define the specification. Therefore,
to guarantee the specifications it is necessary to take compensation data for the
combination of mainframe and modules.
For calibration, you need the dedicated equipment and accessories. Prepare the all
required items or contact your nearest Keysight Technologies service center.