Keysight E5260/E5270 User’s Guide, Edition 6 3-11
Installation
Inspection and Installation
Removal
1. Turn off the mainframe, then wait at least 10 seconds. See Figure 3-2.
2. Disconnect the power cable from rear panel.
3. Disconnect all cables from the modules.
4. Loosen the screws on the plate.
5. Remove the plate.
6. Set the module extractor to the knob position. The extractors are located at the
left and right on the module front panel
7. Hold the module extractor and gently pull out the module.
Installation
1. Align the module with the left and right slot guide rails in the mainframe.
2. Hold the extractor at the left and right on the module front panel.
3. Gently push the module into the slot until you feel it contacts the connector of
the slot.
4. Set the module extractor to the lock position.
5. Replace the plate by using the screws.
6. Check the operation of the mainframe with the modules as follows.
a. Connect the power cable and turn on the mainframe.
b. After the self-test is completed, check the front panel LCD. No mark must be
set to the Error indicator.

Table of Contents

Other manuals for Keysight Technologies E5260 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E5260 Series and is the answer not in the manual?

Keysight Technologies E5260 Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelE5260 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals