7-12 Keysight E5260/E5270 User’s Guide, Edition 6
If You Have a Problem
Error Codes
126 Pulse base and peak must be same polarity.
The polarity of the base and peak values must be the same in the PI
command. Also the polarity of the base, start, and stop values must be
the same in the PWI command.
130 Start and stop must be same polarity.
For a log sweep, the polarity of the start and stop values must be the
same in the WV, WI, WSV, WSI, or WNX command. Also, 0 is not
allowed for the start and stop values.
150 Command input buffer is full.
The Keysight E5260/E5270 can receive 256 characters maximum
including the terminator at one time.
152 Cannot use failed module.
The channel number specifying the module failed the self-test or
calibration. Specify another module that passed the self-test or
calibration. For the service purpose, execute the RCV command to
enable the module.
153 No module for the specified channel.
Module is not installed in the slot specified by the channel number.
160 Incorrect ST execution.
The internal memory programming can be started by the ST command
and completed by the END command. Do not enter the ST command
between the ST command and the END command.
161 Incorrect END execution.
The internal memory programming can be started by the ST command
and completed by the END command. Do not send the END command
before starting the programming.
162 Incorrect command for program memory.
Specified command cannot be stored in the program memory. For the
incorrect commands, refer to Table 2-1 in Programming Guide.