Magtrol Model DSP7000 Dynamometer ControllerChapter 8 – Optional Equipment
operation
8.1.4 i/o CarD 1/CarD 2 CoMManD SeT
Command Code
I/O Card 1
Command Code
I/O Card 2
Function Explanation
IO1AI,# IO2AI,# Read to IO card
Analog input
Analog Channel
1 is available
when not
assigned as the
TACH input.
Read the value of
the AD channel
of the IO card.
Scale factor
unit per volt will
be applied to
reading.
#
1 = Channel 1
2 = Channel 2
Returns a oat
and <cr>< lf>
IO1AO,#,xx.xxx IO2AO,#,xx.xxx Write to IO card
analog output
Analog outputs
are available
when not
assigned to
speed and
torque.
Write the value to
the DA channel
of the IO card.
Volts is assumed.
#
1 = Channel 1
2 = Channel 2
xx.xx
oating point
voltage value 3
decimal places
+10.000 to
-10.000
IO1RL,#,& IO2RL,#,& Write to IO card
Relay output
Relay output 1 is
available when
not assigned to
the alarm output.
Write the value to
the relay.
#
1 = Relay 1
2 = Relay 2
&
0 = Relay off
1 = Relay on
IO1DO,#,& IO2DO,#,& Write to IO card
Digital output
Open collector
NPN
Write the value to
the relay.
#
1 = Output line 1
2 = Output line 2
&
0 = transistor off
1 = transistor on