56
Magtrol Model DSP7000 Dynamometer ControllerChapter 5 – Alarm System
SETUP • OPERATION
5.10 TEMPERATURE ALARM (WB/PB ONLY)
• Toalertuserwhendynamometergetstoohotandthermalswitchopens
• OnlyavailableforusewithWBorPBdynamometers
• Default-alwaysactive
5.10.1 inSTruCTionS for TeMperaTure alarM SeTup
No setup needed.
5.10.2 TeMperaTure alarM aCTion
When the dynamometer in use becomes too hot, the display will flash “TEMPERATURE ALARM
TSCx”. The alarm relay will open and the excitation current will immediately decrease to 10%. After
approximately 3 seconds, the current will drop to 0. Standby signal is cycled for approximately one
second after resetting of the alarm.
POWER
BRAKE STATUS SET POINT SET POINT PID
TORQUE SPEED USER DISPLAY
**** TEMPERATURE ALARM TSCX ****
TORQUE=XXX.XX XX
RPM=X XXXX
Figure 5–23 Temperature Alarm Message Display
5.10.3 To reSeT TeMperaTure alarM
Press any front panel button other than SHIFT. The alarm condition must be cleared before the unit
will return to normal operation.
5.11 ELECTRICAL ALARM
• UsedtoprotecttheDESsupply
• Monitorselectricalinput(mains)andcircuitryoftheDES
• OnlyavailableforusewithWBorPBdynamometers
• Default-alwaysactive
5.11.1 inSTruCTionS for eleCTriCal alarM SeTup
No setup needed.
5.11.2 eleCTriCal alarM aCTion
When there is an electrical overload, the display will flash “ELECTRICAL ALARM TSCx”. The
alarm relay will open and the excitation current will immediately decrease to zero. Standby signal
is cycled for approximately one second after resetting of the alarm.