Magtrol Model DSP7000 Dynamometer Controller
Chapter 7 – Computer Controlled Operation
OPERATION
7.4.1 alarM CoMManDS
Command
Code
Channel 1-
Command
Code
Channel 2
Function Explanation
ALA1,# ALA2,# Enables or disables air ow
alarm.
Values for # are:
0 = disable
1 = enable
ALC,# ALC,# Enables or disables clutch
alarm.
Values for # are:
0 = disable
1 = enable
ALE1,# ALE2,# Enables or disables external
alarm.
Values for # are:
0 = disable
1 = enable
ALL1,# ALL2,# Enables or disables all alarms. Values for # are:
0 = disable
1 = enable
ALP1,xx.xx ALP2,xx.xx Sets power alarm. Sets maximum power in
kilowatts. Range is 0 to
99,999. Setting is applied to
current channel.
ALS1,xx.xx ALS2,xx.xx Sets speed alarm point. Sets maximum speed in rpm.
Range is 0 to 99,999. Setting
is applied to current channel.
ALT1,xx.xx ALT2,xx.xx Sets torque alarm point. Sets maximum torque in input
units. Range is 0 to 10,000.
Setting is applied to current
channel.
ALW1,# ALW2,# Enables or disables water ow
alarm.
Values for # are:
0 = disable
1 = enable
ALR1,# ALR2,# Set up the Relay 1 on the IO
card, either channel one or two,
to be used in conjunction with
that channels alarm system.
See section 5.1.1 Alarm Relay
Values for # are:
0 = disable
1 = enable
GP, xx.xx Set up the global power alarm
on WB/WB or PB/PB tandem
conguration.
Set maximum global power in
Kilowatts. Range is from 0 to
99,999..
GT, xx.xx Set up the global torque alarm
on WB/WB or PB/PB tandem
conguration.
Set maximum global torque
in N·m. Range is from 0 to
99,999..