74
Magtrol Model DSP7000 Dynamometer ControllerChapter 7 – Computer Controlled Operation
OPERATION
7.3.2 TiMeouT
Set the timeout for at least one second if asked to set a communication fault delay timeout.
Note: If the communication fault delay timeout is too short, or if the
computer resets the interface too quickly, the host instrument may
stop responding.
7.4 DSP7000 COMMAND SET
When entering a command code:
1. Type all characters in uppercase ASCII format.
2. End all commands with a CR-LF (hex 0D-0A).
3. Do not string multiple commands together in one line.
The character # represents a floating-point numerical value following the command. Leading zeroes
are not required.
Note: If a command is not recognized, the I/O ERROR message will
appear in the Status Display accompanied by a beep.