62
Magtrol Model DSP7000 Dynamometer ControllerChapter 6 – Manually Controlled Operation
OPERATION
5. After the preload function has been enabled, and the preload value has been set, the set
point the open loop will be initialized to 0.00 % and preload settings is saved to non-volatile
memory.
6. If Scale D button is pressed and the preload function is disabled, the last preload percentage
value will be displayed in set point.
6.7 HOW TO SET AND RESET TARE FUNCTION
6.7.1 SeTTing THe Tare funCTion
1. Press SHIFT button.
2. Press TARE button.
3. “*” indicator will be displayed as in Figure 6–6 Tare function enabled.
POWER
BRAKE STATUS SET POINT SET POINT PID
TORQUE SPEED USER DISPLAY
0.000 W
OFF
0.000 OZ.IN
0000
0%
0% 0%
0
*
Figure 6–6 Tare function enabled
Note: If the tare value is greater than 10% of full scale, the tare function
cannot be enabled.
6.7.2 reSeTTing THe Tare funCTion
1. Press SHIFT button.
2. Press RESET TARE button.
3. “*” indicator will disappear.