Magtrol Model DSP7000 Dynamometer ControllerChapter 7 – Computer Controlled Operation
OPERATION
7.1.2 CHeCking THe DSp7000-To-pC ConneCTion for gpiB SeTup
Note: Make sure that the DSP7000 and its host computer are communicating
before acquiring data.
1. Make sure the primary GPIB address is set correctly for the DSP7000.
2. Set the input variable to 15 characters (13 variable characters and the two required data
termination characters CR and LF. See Section 7.3 – Programming.)
3. Issue output data command “OD” and read 15 characters according to the instructions for
your GPIB interface or serial.
Desired Results
• Torque/speeddatawillbereturned.
• TheerrormessageI/OERRORdoesnotappearonthedisplaypanel.
Note: If the desired results did not occur, please see Chapter 11 –
Troubleshooting.