30
Magtrol Model DSP7000 Dynamometer ControllerChapter 3 – Installation/Configuration
SETUP
3.2.8.2 Software Conguration
1. Turn on the DSP7000. Set up TSC1 as described in section 3.2.5 Eddy-Current or Powder
Brake Dynamometer Setup. Press TSC to swtich to TSC2 setup and proceed to the
dynamometer configuration menu. See Section 3.2.1 – Dynamometer Configuration Menu.
2. Press POWER UNITS until WB or PB is reached.
3. Press MAX SPEED button and use and buttons and Decrease/Increase dial to set
desired scale factor for TSC2.
4. Press SCALE I to add a filter if desired.
5.a. If using an Eddy Current Dynamometer, press SHIFT. The display should appear as shown
in Figure 3–13 TSC2 Eddy-Current Setup Menu. Press the TORQUE UNITS button and use
and buttons and Decrease/Increase dial to set desired max torque. Press the SCALE
P button and use and buttons and Decrease/Increase dial to set desired nominal speed.
Press SHIFT 2 times to return to the setup menu. See Figure 3-21 Tandem Setup Menu.
5b. If using a Powder Brake Dynamometer, press SHIFT. The display should appear as shown
in Figure 3-14 TSC2 Powder Brake Setup Menu. Press the TORQUE UNITS button and
use and buttons and Decrease/Increase dial to set desired max torque. Press SHIFT 2
times to return to the setup menu. See Figure 3-21 Tandem Setup Menu.
6. Press POWER UNITS to select tandem. See Figure 3-22 Tandem Configuration Menu.
7. Press the SCALE I button until the menu says “YES”.
8. Press SHIFT 2 times to complete the initial setup and return to the main menu.