To run the NLDFT report:
1. Go to File > Open. Select a sample file with a Complete status, then click Open.
2. In the view selector drop-down list at the bottom of the window, select NLDFT Advanced
PSD. Graphs for the first sample file display and the sample description shows as the first
group box title in the upper right corner of the window.
A. First selected sample file
B. Second selected sample file
a. Select the Geometry and Model from the drop-down lists for the first sample file.
b. To select isotherm data points for calculation for the first sample file, ensure the option
to the left of the first sample file description is selected. Slide the two blue bars on the
isotherm graph to select data points. Without a second sample selected, the report will
perform a single model DFT calculation and show the results in the two left-hand result
windows.
3. To calculate data from the second sample file, click Select to locate and open the second
sample file with a Complete status. Graphs for the second sample file display and the
sample description will display as the second group box title in the upper right corner of the
window.
a. Select the Geometry and Model from the drop-down lists for the second sample file.
b. To select isotherm data points for calculation for the second sample file, ensure the
option to the left of the second sample file description is selected. Slide the two blue
bars on the isotherm graph to select data points. Data are automatically calculated for
both sample files.
c. Click Edit to make any necessary modifications to the second sample file.
NLDFT Advanced PSD Report