EasyManua.ls Logo

Micromeritics TRISTAR II PLUS - Troubleshooting Valve Operation; Troubleshooting Noisy Vacuum Pump

Micromeritics TRISTAR II PLUS
310 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
11 Troubleshooting
Cause B: Gas may be contaminated.
Action B:
n
Perform a blank analysis. If results are good, perform a reference material
analysis.
n
Replace tank.
n
Check for line leak, which could cause contamination.
n
Flush the lines occasionally to help prevent contamination.
Cause C: Incorrect type of gas line.
Action C: Ensure the gas line is all metal. It is best to use the one shipped with the ana-
lyzer. Do not use polymer gas lines or flexible gas lines that may be coated
internally with a polymer.
Valves cannot be operated.
Cause: Cable from computer to the instrument is loose.
Action: Reconnect the cable.
Vacuum pump is noisy.
CauseA: Sample tube connector is loose.
Action A: Tighten fitting. Replace O-ring.
CauseB: Sample tube O-ring is worn or cracked.
Action B: Replace O-ring. See Sample Tube O-ring Replacement on page11 - 19.
CauseC: Sample tube is cracked.
Action C: Replace with new sample tube.
CauseD: No sample tube loaded on a selected port.
Action D: Install plug or empty sample tube.
CauseE: Gas inlet valve open while vacuum valve open.
Action E: With manual control enabled, use the instrument schematic to close gas inlet
valve.
11 - 2
TriStar II Plus Operator Manual
303-42800-01 (Rev M ) Sep 2023

Table of Contents

Related product manuals