4 Parameter Files
Selections Description
Values used to determine how close the actual pressure must be
to each target pressure from the pressure table. At lower
pressures, the relative tolerance value is lower. At higher
pressures, the absolute tolerance value is lower.
Experiment 1. There might be an absolute tolerance of 5mmHg,
a relative tolerance of 5%, and a target pressure of 40mmHg; 5%
of 40mmHg is 2mmHg. Since 2mmHg (relative tolerance) is
lower than 5mmHg (absolute tolerance), 2mmHg is used.
Therefore a minimum pressure of 38mmHg (40-2) must be
attained to collect data for a target pressure of 40mmHg.
Experiment 2. There might be an absolute tolerance of5 mmHg,
a relative tolerance of 5%, and a target pressure of 200mmHg;
5% of 200mmHg is 10mmHg. Since 5mmHg (absolute
tolerance) is lower than 10mmHg (relative tolerance), 5mmHg is
used. Therefore a minimum pressure of 195 mmHg (200-5) must
be attained to collect data for a target pressure of 200mmHg.
Normally, surface area measurement points are widely spaced
and the resulting measurement is not very sensitive to the precise
location of points so wider tolerances may be used.
Unnecessarily tight tolerances lengthen the analysis.
p° and T [button]
Click to select one option for obtaining the saturation pressure (Po)
and analysis bath temperature. Each selected option presents a
different set of parameters at the bottom of the window.
Options and descriptions:
1. Analysis bath temperature (not used for krypton analyses).
Measures the p
0
on a continuous basis. Allows the meas-
urement of each data point without slowing the analysis.
Analysis Conditions (continued)
4 - 10
TriStar II Plus Operator Manual
303-42800-01 (Rev M ) — Sep 2023