4 Parameter Files
Selections
Description
Copy QuickStart
degas conditions
from SmartVacPrep
Unit [drop-downbox]
Use to copy the degas conditions settings from the selected
SmartVacPrep unit and port.
Degas conditions
[drop-downbox]
Use to browse for a .DEG file that contains degas condition para-
meters to be used in the analysis.
Heating Phase [table]
This option is applicable when degassing with a SmartVacPrep.
Enter up to five stages of degas conditions.
Temperature. Temperature at which the sample is to be held while
degassing.
Time. How long the sample is to be held at the specified temperature
before beginning to cool down.
Temperature Ramp Rate. The rate at which the temperature will
change while advancing to the hold temperature.
Degas Conditions
4 - 14
TriStar II Plus Operator Manual
303-42800-01 (Rev M ) — Sep 2023