A5/A3™ Operating Instructions 046-003777-00 4 - 15
Preoperative Tests Leak and Compliance Tests
3. Compare the test results with the information in TABLE4-4, “Manual Circuit Leak Test
Results,” on page4-15, and proceed accordingly.
RESULTS COMMENTS/OPTIONS
Pass
Example:
Manual Circuit Leak Test passed.
Adjust the APL valve to SP position.
Select Continue to proceed to the main screen.
Fail: Fresh gas flow detected
Example:
Manual Circuit Leak Test failed.
Fresh gas is detected.
Adjust all flowmeters to zero.
Select Retry to repeat the test.
TABLE 4-4 Manual Circuit Leak Test Results