EasyManuals Logo

Mindray A5 Operating Instructions

Mindray A5
310 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #213 background imageLoading...
Page #213 background image
A5/A3â„¢ Operating Instructions 046-003777-00 7 - 23
Maintenance Cleaning and Disinfection
Gasket
4. Cleaning
a. Clean the absorber canister with a soft, lint-free cloth and a recommended cleaning
agent (see section7.10.2 (page 7-11) "Cleaning and Disinfecting Agents / Autoclaving").
Allow them to dry thoroughly.
b. If disinfecting the absorber canister, continue with step 5, otherwise skip to step 6.
5. Disinfection
NOTE: Ensure that the absorber canister has been cleaned as described in step
4 before disinfecting.
See section7.10.2 (page 7-11) "Cleaning and Disinfecting Agents / Autoclaving" and use an
approved disinfecting agent for the absorber canister while adhering to facility policies and
procedures.
6. Make sure that the gasket is correctly installed. The comparison between correct installation and
incorrect installation is shown below.
FIGURE 7-19 Absorber Canister, Unlocked FIGURE 7-20 Absorber Canister, Locked
Unlocked Position
Gasket
Locked Position

Table of Contents

Other manuals for Mindray A5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A5 and is the answer not in the manual?

Mindray A5 Specifications

General IconGeneral
TypeAnesthesia Machine
BrandMindray
ModelA5
Power Supply100-240 V AC, 50/60 Hz
Ventilation ModesSIMV, PSV
FlowmetersO2, N2O, Air
Vaporizer CapacityUp to 3 vaporizers
Safety FeaturesO2 failure protection, low pressure alarm
Gas ModuleOptional integrated gas module
Integrated Air CompressorOptional
Patient MonitoringIntegrated monitoring with optional modules

Related product manuals