No. 99MBB091A
5 - 16
■ Procedure for setting ls to None
The SJ-301 allows the s filter to be set to None, only if the roughness standard has been set
to “JIS1994” and the measured profile has been set to “R profile”.
NOTE If the roughness standard is other than “JIS1994” and if the measured profile is other
than “R profile”, the setup and settings are automatically determined.
● Operating procedure
Condition
JIS1994
GAUSS
c 0.25mm 5
(AUTO)
RANGE
Ra
Ry
CAL.
* * * *
* * * *
2004-05-13
m
m
CARD
F1 F2 F3 F4
N5
Condition menu
STAND.
PROFILE
FILTER
c
s
EVA-L
RE-CAL.
JIS'94
R
GAUSS
4.0mm
0.8mm
2.5m
Page1/3
N5
Condition menu
STAND.
PROFILE
FILTER
c
s
EVA-L
RE-CAL.
JIS'94
R
GAUSS
4.0mm
0.8mm
2.5m
Page1/3
s
2.50.8
25
None
8
0.25
1
<Home screen>
[Measurement conditions]
2
<Measurement Condition Setup screen>
3
<s Setup window>
s Setup buttons
[s]
Touch the [Measurement Conditions] button in the Home
screen.
The Measurement Condition Setup screen will appear.
Touch the [s] button.
The s Setup window will appear.
Touch one of the buttons to be set on the s Setup window.
The touched setup button is highlighted.
None: The s filter is set to None.