6 - 1
No. 99MBB091A
RECALCULATION WITH
THE MODIFIED
MEASUREMENT
CONDITIONS
The SJ-301 has a function to recalculate the measurement data by modifying the measure-
ment conditions after performing a roughness measurement. If this recalculation function is
on, measurement data is recalculated with the modified measurement conditions and dis-
played on the touch panel.
Measurement conditions that can be modified
The SJ-301 can perform recalculation after modifying the following measurement conditions:
Surface texture standard
Measured profile
Filter
Number of sampling lengths (can be reduced)
Parameters
Unit
Deletion of the required data
NOTE This recalculation function cannot be used if MOTIF.R-profile or MOTIF.W-profile
is specified.
The recalculation function cannot be used if the number of sampling length is
increased.
The SJ-301 does not support recalculation with a modified cutoff length.
The recalculation function may not be used due to mismatch of the pre-travel and
post-travel conditions if the filter or measured profile is modified.
Transition of screens if recalculation is performed
Home screen
(Execution of measurement)
Measurement Condition Setup screen
(Modification of measurement conditions)
(Execution of recalculation)
Home screen
(Confirmation of
recalculation result)
[Condition]
[Home]
[RE-CAL.]
Condition
JIS1994
GAUSS
c 0.8mm 5
(AUTO)
RANGE
Ra
Ry
CAL.
1.54
7.41
2000-01-01
m
m
CARD
Condition
ISO1997
2CR
c 0.8mm 3
(AUTO)
RANGE
Ra
Ry
CAL.
1.55
7.44
2000-01-01
m
m
CARD
Condition menu
STAND
PROFILE
FILTER
c
s
N
EVA-L
RE-CAL.
ISO'97
R
2CR
2.4mm
3
0.8mm
2.5
m
Page1/2
F1 F2 F3 F4 F1 F2 F3 F4