5.10.7 Setting the calculation conditions when selecting a profile motif (MOTIF.R/
MOTIF.W)
The SJ-301 can use either ISO 12085 Main Body compatible method or ISO 12085 Annex A
compatible method for the motif connection method when selecting a profile motif
(MOTIF.R/MOTIF.W).
TIP Setting-up of the motif connection method is made in the Parameter Select screen.If
other than a profile motif is selected, the connection method setup is invalidated.
Annex A
Annex A
: ISO 12085 Standard Annex A compatible connection method
: ISO 12085 Standard Body compatible connection method
Screens displayed when setting the motif connection method
Annex A
Select parameter
Custom
Wx
Wte
P2
P3
Rx W
AW
P1 R AR
P4
ESC
MOTIF.R
P5
s
B
EVA-L
Condition menu
Page1/3
STAND.
L
RE-CAL.
ISO'97
MOTIF.R
A
0.5mm
16.0mm 16.0mm
8m
PROFILE
FILTER
Home screen
Measurement Condition
Setup screen (1/3)
Measurement Condition
Setup screen (2/3)
Parameter Select screen
[Page2/3]
[Page1/3]
[Home]
[Home]
[Cstomize]
[Enter]
[Measurement
conditions]
Condition
JIS1994
GAUSS
0.5mm 5
(AUTO)
RANGE
R
AR
CAL.
* * * *
* * * *
2004-05-15
m
m
CARD
F1 F2 F3 F4
0.5mm/s
OFF
Page2/3
GO/NG
mm/inch
CUST.
mm
RANGE AUTO,ESC
RE-CAL.
Condition menu