14 - 15
No. 99MBB091A
14. REFERENCE INFORMATION
14.4 Traversing Length
A traversing length is defined as being the evaluation length plus the approach travel length,
pre-travel length, and post-travel length.
NOTE The pre-travel length and the post-travel length will vary depending on the filter used.
If the pre-travel and post-travel setting is set to “No”, the traversing length is reduced
by as much as the pre-travel length and the post-travel length.
Refer to “5.9 Setting Pre-Travel/Post-Travel to “OFF” ”.
Measuring operation
1 cycle reciprocal operation
0.5mm/s (.02in/s)
1mm/s (.04in/s)
Measurement starts from the origin limit position. If measurement has been completed, the
detector returns to the origin limit.
Traversing Length
If the 2RC filter is selected
Approach
travel
0.5mm
(.02")
ᐉᐉ×n
( = λc)
Pre-travel
Evaluation length
Origin
Normal traversing
end
Traversing length
If the GAUSS filter is selected
0.5mm
(.02")
/2
Approach
travel
/2 ×n
Pre-travel
( = λc)
Evaluation length
Origin
Normal traversing
end
Traversing length
Post-travel
Data from the pre-travel and post-travel lengths are calculated assuming that their lengths are
/2.