No. 99MBB091A
8 - 2
8.1 Data To Be Saved and Storage Media
■ Objective save/recall data and its media
Data saving and recalling is outlined below, where data is divided into three groups according
to data handling. When each data group is saved, the print conditions and calibration condi-
tions are saved as well.
Data group Stored contents Storage media
Save
Buttons and screens used for operation
Measurement
conditions
Measurement
conditions,
Print conditions,
DATA key function
setup,
Calibration
conditions
Internal memory
(5 files max.), or
Memory card,
(20 files max.).
Note: Only recall from the internal
memory is permitted.
Measured
profile
Measurements,
Profile data points,
Measurement
conditions,
Print conditions,
DATA key function
setup,
Calibration
conditions
Memory card only
Statistical data Statistical results,
Profile data points
Measurement
conditions,
Print conditions,
DATA key function
setup,
Calibration
conditions
Memory card only
Recall
START
STOP
PRINT
DATAFEED
CONDITION
READ
[CONDITION READ] key
[CONDITION READ] key
on the Display Unit
Note: DATA key function setup
required.
START
STOP
PRINT
DATAFEED
CONDITION
READ
[DATA] key
[DATA] key on the Display Unit
Home screen
Condition
JIS1994
GAUSS
c 0.8mm 5
(AUTO)
RANGE
Ra
Ry
CAL.
1.54
7.41
2000-01-01
m
m
CARD
Measurement Condition
Setup screen (1/3)
Measurement Condition
Setup screen (1/3)
Statistical Processing
Results screen
Statistics
Ra
1999/01/15
SAMPLE NUM
(050)
0.02 m
3.01 m
3.25 m
2.70 m
100.0%
X
MAX
MIN
PASS RATIO
ESC
Statistical Condition
Setup screen
Statistics
Ra
MIN
2.00
MAX
4.00
50
0.2
GO/NG
PARA.
INTER.
TOL.
NUM
ESC
[Data save/recall]
Home screen
Condition
JIS1994
GAUSS
c 0.8mm 5
(AUTO)
RANGE
Ra
Ry
CAL.
1.54
7.41
2000-01-01
m
m
CARD
[Data save/recall]
[Save] [Recall]
F1 F2 F3 F4 F1 F2 F3 F4
N5
Condition menu
STAND.
PROFILE
FILTER
c
s
EVA-L
RE-CAL.
JIS'94
R
GAUSS
4.0mm
0.8mm
2.5m
Page1/3
N5
Condition menu
STAND.
PROFILE
FILTER
c
s
EVA-L
RE-CAL.
JIS'94
R
GAUSS
4.0mm
0.8mm
2.5m
Page1/3
[Save/Recall] [Save/Recall]
NOTE When an objective data item is read, the existing SJ-301 main unit setup will be
overwritten with the above described “storage contents” that are read together.