EasyManua.ls Logo

Mitutoyo SJ-301 - Page 117

Mitutoyo SJ-301
346 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5 - 19
No. 99MBB091A
5. MODIFYING THE MEASUREMENT CONDITIONS
Modifying the number of sampling length
Condition
JIS1994
GAUSS
c 0.8mm 5
(AUTO)
RANGE
Ra
Ry
CAL.
1.54
7.41
2000-01-01
m
m
CARD
F1 F2 F3 F4
N5
Condition menu
STAND.
PROFILE
FILTER
c
s
EVA-L
RE-CAL.
JIS'94
R
GAUSS
4.0mm
0.8mm
2.5m
Page1/3
N5
Condition menu
STAND.
PROFILE
FILTER
c
s
EVA-L
RE-CAL.
JIS'94
R
GAUSS
4.0mm
0.8mm
2.5m
Page1/3
N
15
L 0.3mm
3
N5
Condition menu
STAND.
PROFILE
FILTER
c
s
EVA-L
RE-CAL.
JIS'94
R
GAUSS
4.0mm
0.8mm
2.5m
Page1/3
N
53
L 0.3mm
1
<Measurement Condition Setup screen>
[N] (the number of sampling lengths)
[Condition]
2
1
<Home screen>
3
<Number (N) Setup window>
Number (N) button
4
<Number (N) Setup window>
Title bar
Touch the [Condition] button on the Home screen.
The Measurement Condition Setup screen will appear.
Touch the [N] button.
The Number (N) Setup window will appear.
Touch the button of the desired number.
The touched button will be highlighted.
TIP If the [L] button is touched, an arbitrary evalua-
tion length can be set. Refer to section, 5.6
“Setting the Evaluation Length to an Arbitrary
Length” for details.
Touch the title bar of the Number (N) Setup window to accept
the selected number of sampling lengths.
The Number (N) Setup window closes.

Table of Contents