5. MODIFYING THE MEASUREMENT CONDITIONS
12
<Traversing Speed Setup window>
11
<Calibration Condition Setup Screen>
[SPEED]
9
<Number (N) Setup window>
10
<Number (N) Setup window>
Title bar
Calibration
CONDITION
STD
ESC
Calibration
Condition
STD
ESC
Calibration
Condition
STD
ESC
0.5mm/s
c
FILTER
GAUSS
N3
0.8mm
0.5mm/s
c
FILTER
GAUSS
N5
2.5mm
0.5mm/s
c
FILTER
GAUSS
N5
2.5mm
N
5
31
Calibration
Condition
STD
ESC
0.5mm/s
c
FILTER
GAUSS
N5
2.5mm
N
3
1
5
0.25
0.5
Sampling length number button
Traversing speed button
Touch the desired number button to specify the number of
sampling lengths.
The button touched will be highlighted.
Touch the title bar in the Number (N) Setup window.
The Number (N) Setup window will be closed.
Touch the [Speed] button.
The Traversing Speed Setup window will appear.
Touch the traversing speed button to be set.
The touched traversing speed button will be highlighted.
NOTE If the cutoff length (c) is set to either of the
following values, the traversing speed is fixed.
If the cutoff length is 0.08 mm:
Traversing speed is 0.25 mm/s.
If the cutoff length is 0.25 mm:
Traversing speed is 0.25 mm/s.
If the cutoff length is 2.5 mm:
Traversing speed is 0. 5 mm/s.