11. MAINTENANCE AND INSPECTION OF SJ-301
6
5453 5655 5857
4645 4847 5049 5251
3837 4039 4241 4443
3029 3231 3433 3635
2221 2423 2625 2827
1413 1615 1817 2019
65 87 109 1211
LCD/SW 21 43
<Touch Panel Test screen>
5
5453 5655 5857
4645 4847 5049 5251
3837 4039 4241 4443
3029 3231 3433 3635
2221 2423 2625 2827
1413 1615 1817 2019
65 87 109 1211
LCD/SW 21 43
<Touch Panel Test screen>
[LCD/SW]
<LCD Test screen (test in progress)>
Test buttons
(numbered)
Touch the test buttons one by one, starting with the [1] button
to highlight all the buttons.
If the button touched is highlighted, the LCD panel is
normal for the area that corresponds to the touched button.
If it is not highlighted, the LCD panel has something
wrong with touch function for the area that corresponds to
the touched button.
NOTE Do not touch the [LCD/SW] button during this
LCD button test.
If the [LCD/SW] button is touched another test
will be initiated.
Touch the [LCD/SW] button after all the test buttons have
been touched.
The LCD Test screen (test in progress) will be gradually
filled from the top.
If the entire screen is filled in, the LCD has been found to
be normal. If any sections are not filled, the display
function of the LCD may not be operating normally.
After the entire screen has been filled, a blank (back-
ground color) display will rise from the bottom section.
After a short period the original Touch Panel Test screen
will be restored.