14 - 11
No. 99MBB091A
14. REFERENCE INFORMATION
Phase correct filters
Output waveforms from the general 2RC filters may be distorted due to phase deviations
that vary with each wavelength.
Shown below are the response of a low-pass filter to square wave input, and that of a high-
pass filter.
Input signal (square wave)
500.00 μm/cm(x 20)
0.20 μm/cm(x 50000)
LPF: PC
500.00 μm/cm(x 20)
0.20 μm/cm(x 50000)
LPF: 2RC
500.00 μm/cm(x 20)
0.20 μm/cm(x 50000)
Low-pass filter
HPF : PC
500.00 μm/cm(x 20) 500.00 μm/cm(x 20)
0.50 μm/cm(x 20000)
HPF : 2RC
0.50 μm/cm(x 20000)
High-pass filter