Running an Experiment
192 0112-0109 H
Experimental Setup Parameter
Fluid Transfer Settings—Aspirate
Parameter Setting
Read Mode Luminescence
Excitation Wavelength None
Emission Wavelength None
Camera Gain
1
1. May be adjusted to compensate for cell brightness or saturation problems.
70,000–280,000
Exposure Time
1
0.53 s
LED Excitation Intensity NA
Gate Open 100%
Read Interval 1 s
Reads before pipetting 5
Reads during first interval 50–90
Save Assay Images Possible, but creates very large files
Parameter
Suspension Setting
(pipette cells)
Adherent Settings
(pipette compound)
Cell Flask Spinning Rate Manual setting, Speed 6 NA
Fluid Transfer Type Single Aspirate, Single
Dispense
Single Aspirate, Single
Dispense
Source Plate
Fill Reservoir speed
Drain Reservoir speed
Drain destination
defaultcellres384
5
4
Cell flask
Source Plate 2
NA
NA
NA
Aspiration Volume 25 μL 25 μL
Aspiration Height 70 μL5 μL
Aspiration Speed 20 μL/s 20 μL/s
Tip Up Speed 10 mm/s 10 mm/s
Hold Volume
1
0 μL 0 μL

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Molecular Devices FLIPR Tetra and is the answer not in the manual?

Molecular Devices FLIPR Tetra Specifications

General IconGeneral
BrandMolecular Devices
ModelFLIPR Tetra
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals