FLIPR
®
Tetra High Throughput Cellular Screening System User Guide
0112-0109 H 201
103: “Invalid start of
command.”
Start of command is
not valid. See
appended string on
the error message for
details.
Contact Technical Support
at
+1-800-635-5577.
104: “Command invalid,
system busy.”
The system cannot
act upon the
command, as the
system is busy. See
appended string on
the error message for
details.
Contact Technical Support
at
+1-800-635-5577.
“Data buffer full”. The data buffer is full.
See appended string
on the error message
for details.
Contact Technical Support
at
+1-800-635-5577.
106: “Option not installed,
command unavailable”.
Pipettor head, LED,
Emission Filter or
TETRAcycler™ is not
installed.
Install option you want to
use.
Change protocol to use
available instrument
options.
107: “Command failed”. General failure. Contact Technical Support
at
+1-800-635-5577.
108: “Command failed,
system in an error state".
Minor communication
error occurred.
Select Clear Error
command from
Instrument drop-down
menu.
109: “Command failed,
system not initialized “.
Communication failed
between software and
instrument.
Select Reset from the
Instrument drop-down
menu.
Turn the instrument
power switch OFF then ON
to cycle power.
“Password does not
match”.
User password is
incorrect.
Contact your instrument
administrator for the
appropriate user name
and password.
Table 9-3 General Troubleshooting (cont’d)
Symptom or Error
Message
Possible Causes Solutions