MF motor off
IMM PGM MIP
USAGE ♦ ♦ ♦
Units xx - none
1 - motor power is ON
ST - stop motion
SYNTAX xxMF or xxMF?
PARAMETERS
Description xx [ int ] - axis number
Range xx - 1 to MAX AXES
Defaults xx missing: error 37, AXIS NUMBER MISSING
out of range: error 9, AXIS NUMBER OUT OF RANGE
DESCRIPTION This command turns power OFF of the specified motor (axis).
RETURNS If “?” sign is issued along with command, the controller returns:
0 - motor power is OFF
REL. COMMANDS AB - abort motion
MO -
0 | controller returns a value of 0
urn motor power ON
EXAMPLE 2MF | turn axis #2 motor power OFF
2MF? | query axis #2 motor power status
2MO | turn axis #2 motor power ON
2MF? | query axis #2 motor power status
1 | controller returns a value of 1
Section 3 – Remote Mode 3-97