Specifications
This section lists the technical specifications for the LabChip GX
Touch/GXII Touch instruments. Technical specifications are subject
to change without notice.
General
Environmental
Electrical
Assay Voltage
Chip Pressure
Size (H x W x D) 27” (69cm) x 20” (51cm) x 19” (49cm)
Weight 56 lbs. (25.5 kg)
Plate Capacity (1) 96-well or 384-well microplate
Ventilation/Cooling 3” (77 mm) minimum space required around instrument for
proper air flow.
Operating Temperature 65° to 78°F (18° to 26°C)
Operating Humidity 20% to 70% relative humidity, noncondensing
Storage Temperature 50° to 104°F (10° to 40°C)
Transient Overvoltages Installation Category II Overvoltage
Pollution Pollution Degree 2
Altitude Up to 2000m
Indoor Use Only
AC Input 100-240Vac, 50-60Hz (±10%)
Power Rating 300VA
Fuses (2) 3.15A Slo-Blo
®
250V, 5x20mm (P/N CLS138847)
Grounding Through the power cord
Computer Interface USB 2.0
Minimum Voltage/Current -100V (limited by current)
Maximum Voltage/Current -3000V (limited by current)
Pressure Range +50 to -5 psig

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the PerkinElmer LabChip GXII Touch and is the answer not in the manual?

PerkinElmer LabChip GXII Touch Specifications

General IconGeneral
BrandPerkinElmer
ModelLabChip GXII Touch
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals