EasyManua.ls Logo

PerkinElmer LabChip GXII Touch - Changing the Fuses

PerkinElmer LabChip GXII Touch
190 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Maintenance and Service 142
P/N CLS151164 Rev. B LabChip GX Touch/GXII Touch User Manual PerkinElmer, Inc.
Changing the Fuses
The LabChip GX Touch/GXII Touch instruments use two fuses
located in the power entry module:
(2) 3.15A Slo-Blo
®
250V, 5x20mm (P/N CLS138847)
To order replacement fuses, contact PerkinElmer Technical
Support. See Contact Us on page 3.
WARNING
Electric shock Hazard. Disconnect the power cord before
changing the fuses.
For continued fire protection and correct functioning of the unit,
replace fuses only with the exact part number to prevent fire.
To replace the fuses:
1 Verify the LabChip GX Touch/GXII Touch instrument power is
OFF.
2 Unplug the power cord from the instrument.
3 Use a small flat-blade screwdriver at the top of the fuse holder
to gently pry the fuse holder out of the power entry module (see
Figure 60).
Figure 60. Power Entry Module
4 Remove the two 3A fuses from the fuse holder (see Figure 60).
Figure 61. Fuse Holder and Fuse

Table of Contents

Related product manuals