Maintenance and Service 141
P/N CLS151164 Rev. B LabChip GX Touch/GXII Touch User Manual PerkinElmer, Inc.
Cleaning Test Chip C
Before using Test Chip C, verify the test chip is clean. The chip
must be clean before running Diagnostic tests.
1 Inspect the ruby crystal, chrome mask, and all glass surfaces for
debris and/or fibers that may scatter light, fluoresce, or block
excitation/emission light.
Figure 59. Top and Bottom of Test Chip C
2 Dust can be removed with a gentle air stream from oil free
compressed and filtered house air or compressed air cans. Do
not shake can before use. Hold can at a distance from the chip
to prevent condensation.
3 To remove dyes or reagents from the Ruby crystal, chrome
mask, and glass surfaces:
a Do not apply any force when cleaning.
b Use a lint free cloth and/or swabs (provided in accessory kit)
dampened with 70%-Isopropanol / DI water solution.
c Remove liquid from all surfaces with a dry lint free cloth or
dry swabs.
d Dry the chip completely with a gentle air stream from oil free
compressed and filtered house air or compressed air cans.
Do not shake can before use. Hold can at a distance from
the chip to prevent condensation.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the PerkinElmer LabChip GXII Touch and is the answer not in the manual?

PerkinElmer LabChip GXII Touch Specifications

General IconGeneral
BrandPerkinElmer
ModelLabChip GXII Touch
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals