Preparing the Instrument to Run Samples 25
P/N CLS151164 Rev. B LabChip GX Touch/GXII Touch User Manual PerkinElmer, Inc.
Inserting the Chip
To insert the chip into the instrument:
1 Prepare the chip and reagents for the assay as directed in the
Assay Quick Guide. Follow the instructions carefully to properly
prepare the chip.
2 Touch the Home button on the Navigation Bar.
3 Touch the Unload Chip button on the Home Window. The chip
door on the front of the instrument unlatches.
4 Lift the chip door up.
5 Place the chip into the cutout with the sipper through the hole in
the chip holder.
6 Close the chip door and press down to latch the door.
7 If the chip supports multiple assays, a dialog prompts you to
select the assay that the chip will be used for. Select the desired
assay and touch the OK button.
8 If a Protein Clear HR chip is inserted into the instrument,
calibrate the instrument to run Protein Clear HR assays (see
page 26).
9 If any other chip is inserted into the instrument, the chip can be
primed while preparing the first plate (see page 27).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the PerkinElmer LabChip GXII Touch and is the answer not in the manual?

PerkinElmer LabChip GXII Touch Specifications

General IconGeneral
BrandPerkinElmer
ModelLabChip GXII Touch
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals