Running an Assay 42
P/N CLS151164 Rev. B LabChip GX Touch/GXII Touch User Manual PerkinElmer, Inc.
Washing the Chip
Some protein samples may have components which produce data
with extra peaks, spikes or other artifacts. When these artifacts are
present, washing protein chips on the LabChip GXII Touch
immediately before the next use can often restore data quality.
1 Prepare the chip and buffer tube as instructed in the Assay User
Guide.
2 Insert the prepared chip and buffer tube into the instrument (see
page 25).
3 Touch the Home button on the Navigation Bar.
4 Touch the Wash button on the Home Window.
5 When the wash is complete, follow the instructions in the Assay
User Guide to run the chip.