Viewing the Data 50
P/N CLS151164 Rev. B LabChip GX Touch/GXII Touch User Manual PerkinElmer, Inc.
b Touch each well to view in the overlay graphs. Multiple wells
can be selected at the same time. Touch a well again to
deselect the well.
5 To display the analyzed data on the Egram/Gel Tab, EGram Tab
and Gel Tab, select the Perform Sample Analysis check box.
To display the raw data only, clear the Perform Sample
Analysis check box.
6 Use the gels in the Gel Tab or the Egram/Gel Tab to select the
wells if there are multiple cycles. Touching a well in the Select
Wells to View Tab selects the well data from the last plate cycle.
7 To view the data in the Egram/Gel Tab, see page 50.
8 To view the graphs in the EGram Tab, see page 52.
9 To view the data in the Gel Tab, see page 52.
10 See Zoom In and Zoom Out on page 52 for details on zooming
graph and gel views.
Viewing Data in the Egram/Gel Tab
The Egram/Gel Tab displays a graph at the top of the tab and the
gels of all the well data at the bottom of the window. The graph at
the top of the window can display a single well or multiple wells
overlaid in the same graph. Gels for all wells in the run display at
the bottom of the window.
The wells selected on the Select Wells to View Tab display in the
Egram/Gel Tab.
The wells to view can also be selected using the gels on the bottom
of the Egram/Gel tab:
1 Select or clear the Overlay check box.
2 Touch a gel at the bottom of the tab to add the graph to the
overlay. If the overlay check box is selected, the most recently
added well displays at the top of the overlay graph. If the
Overlay check box is not selected, only one well can be
selected.
3 To change the graph display options, select the desired options
in the Graph Settings on the EGram Tab.
4 To change the gel display options, select the desired options in
the Gel Settings on the Gel Tab.