Software Reference 72
P/N CLS151164 Rev. B LabChip GX Touch/GXII Touch User Manual PerkinElmer, Inc.
Home Window
The Home Window displays to the right of the Navigation Bar and is
used to prime a chip, calibrate a chip, wash a chip, run a plate, or
purge the instrument’s pressure lines. Also used to load and unload
plates and chips in the instrument. The status of the instrument and
user information also display.
If the LabChip GxP option is installed, the Home Window displays
the CDR Server status and provides access to the Login window.
To open the Home window, touch the Home button on the
Navigation Bar.
Figure 19. Home Window
The Home window contains the following buttons:
Prime - Opens the Prime Window to prime the chip (not available for
Protein Clear HR chips).
Prime and Calibrate - Opens the Prime and Calibrate Window to
calibrate the instrument to run Protein Clear HR assays (only for
Protein Clear HR chips).