Running an Assay 38
P/N CLS151164 Rev. B LabChip GX Touch/GXII Touch User Manual PerkinElmer, Inc.
Monitoring the Run (Continued)
NOTE
If the ProteinEXact calibration steps do not display, restart the
software and prime the chip.
5 Data is saved to a GX Touch data file (*.gxd) with the name
shown in the Setup Run Tab after each well is completed.
6 To view the results for individual wells as data is acquired or
after the run is finished, see page 49.
7 If analysis settings are changed during the run (see page 38),
the analyzed data uses the current analysis settings if Perform
Sample Analysis is selected on the Select Wells Tab.
8 To stop the run before it is complete, see page 39.
9 When the assay is complete, Successfully Completed displays
in the Status line on the Status Window.
10 Remove the plate (see page 41) and/or remove the chip (see
page 41) as necessary.
Editing the Analysis Settings during a Run
Specific analysis settings can be adjusted in the LabChip GX Touch
software while a plate is running. If Auto Export is selected on the
Setup Run Tab, the Analysis Settings cannot be changed unless the
Defer Export to Run Completion check box is selected.
After a run is complete, the analysis settings can be changed to
view the data, but the analysis settings cannot be saved.
To adjust the analysis settings:
1 After the run is started, touch the Views button and then the
Analysis Settings button. The Analysis Settings Tab opens.
2 Change the Analysis Settings as desired. Press Enter or Tab, or
move to another text box, to enable the Apply and Cancel
buttons. (See Analysis Settings Tab for a description of each
analysis setting.)
3 Click the Apply button to apply the settings to the current run.
The initial analysis settings are saved as Version 0. If changes
are made during the run, the analysis settings at the end of the
run are saved as Version 1. Use LabChip Reviewer to view the
analysis settings versions.