Preparing the Instrument to Run Samples 27
P/N CLS151164 Rev. B LabChip GX Touch/GXII Touch User Manual PerkinElmer, Inc.
Priming the Chip Before the Run
If a new chip that is not a Protein Clear HR chip has been inserted
into the instrument, the chip is automatically primed at the start of
the first run.
Since priming is a lengthy process, the chip priming can be started
before starting the run. (To save time, the sample plate can be
prepared while the chip is priming.)
To prime a chip before starting the run:
1 Insert the chip (see page 25).
2 Load the buffer tube (see page 31). If running a test ladder, load
the ladder tube at the same time.
3 Touch the Home button on the Navigation Bar.
4 Touch the Prime button on the Home Window. The Prime
Window opens.
5 On the Prime Window, select the assay for which to prime the
chip on the Assay drop-down list. (To select the location of the
assay file, touch the Change Assay Folder button and select
the assay folder on the Change Assay Folder Window.)
6 If desired, select the Run Test Ladder after Prime check box to
run one ladder after the prime is complete.
7 Touch the Prime button. The Prime Window displays the time
left until the prime is complete.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the PerkinElmer LabChip GXII Touch and is the answer not in the manual?

PerkinElmer LabChip GXII Touch Specifications

General IconGeneral
BrandPerkinElmer
ModelLabChip GXII Touch
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals