Preparing the Instrument to Run Samples 30
P/N CLS151164 Rev. B LabChip GX Touch/GXII Touch User Manual PerkinElmer, Inc.
Defining a New Plate Type (Continued)
The instrument will move to the specified number of wells in
each corner of the plate, punching holes in the tape.
The Punch test is acceptable if the punched hole is >0.5mm
from the edge of the well.
Holes may not be perfectly centered.
Possible sources of errors:
Sipper splay
Plate tolerances
Robot alignment error
If all holes are too close to the edge (<0.5mm) in the same
direction, adjust the X-Margin or Y-Margin values on the Add
Plate window.
8 Select the new plate name on the Setup Run Tab when starting
the run.