Running an Assay 34
P/N CLS151164 Rev. B LabChip GX Touch/GXII Touch User Manual PerkinElmer, Inc.
Defining the Run Parameters (Continued)
To select or clear adjacent wells, touch and drag over the
desired wells.
To use a Sample Names file to select the wells, see
Selecting Wells using a Sample Names File on page 57.
7 If desired, select the Column-wise check box to sample the
plate in column order. If not selected, the plate is sampled in row
order.
8 If desired, select the Advanced Settings check box.
To repeatedly run the selected wells and combine the data
into one data file, touch the + or - button under Sample Sips
to select the number of times to sample the selected wells.
To perform the assay multiple times on the same plate,
touch the + or - button under Plate Cycles to select number
of times to run the assay.
To randomly sample a specific percent of the selected wells,
touch the + or - button under Random Select% to specify
the percent of wells to sample during the run.
9 Touch the Next button . The Setup Run Tab opens.
10 If desired, type the name of the operator in the Operator text
box. (If the LabChip GxP option is installed, the name of the
current user is automatically displayed and cannot be edited.)
11 For ProteinEXact chips that have been primed before the run
(see page 27), select the Skip Warm and Skip Prime check
boxes to avoid repeating the priming and warming process.
12 To read the plate barcode (if your instrument is equipped with a
barcode reader), select the Read Plate Barcode check box.
(The plate barcode can be used as part of the data file name if
selected.)
13 To change the Data Path, touch the Browse button, select the
desired location for the data file in the Browse for Folder
window, and touch the OK button. If the LabChip GxP option is
installed, the data file must be located in the CDR. To restore
the default path, touch the Default button.
14 Select the Copy to check box to automatically create a copy of
the data files. Touch the Browse button, select the desired
location for the data files in the Browse for Folder window, and
touch the OK button. To restore the default path, touch the
Default button.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the PerkinElmer LabChip GXII Touch and is the answer not in the manual?

PerkinElmer LabChip GXII Touch Specifications

General IconGeneral
BrandPerkinElmer
ModelLabChip GXII Touch
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals